Biomedical Imaging by Youxin Mao - HTML preview

PLEASE NOTE: This is an HTML preview only and some elements such as links or page numbers may be incorrect.
Download the book in PDF, ePub, Kindle for a complete version.

iam

D

#3

Measured X

Fig. 16 (a) and (d) show SS-OCT images of human finger in vivo taken using fiber probes #

20

Measured Y

eam

14 (working distance, depth of field and spot diameter of 1.2 mm 1.1 mm, and 27 m) and #

#1

Gaussian Fitted X

B

Gaussian Fitted Y

16 (working distance, depth of field and spot diameter of 0.9 mm 0.33 mm, and 16 m)

10

acquired by our catheter-based complex SS-OCT using our 3x3 Mach-Zehnder

0.0

0.5

1.0

1.5

2.0

interferometer with unbalanced differential detection technique with image size of 5x2mm.

Distance to Lens Surface (mm)

The image depth shown in Fig. 16 (a) is slightly larger than that in Fig. 16 (b), but the image

Fig. 15. Measured and Gaussian-fitted 1/e2 intensity beam diameters along the axial

is blurrier in Fig. 16 (a) than that in Fig. 16 (b), which taken by the probe with larger depth of

distance (zero is the position of the lens surface) at x (horizontal) and y (vertical) radial

field and spot size. The image shown in Fig. 16 (b) has higher resolution than that in Fig. 16

coordination in the distance range of depth of field of the samples #1, #3 and #8, which was

(a), which can be seen clearly with finer structures in layer of epidermis (grey arrow), sweat

made from the GRIN fiber lens. Insets: the cross-section beam profiles at three planes for the

gland (white arrow), and blood vessel in subcutis layer (black arrow).

sample #8.

For each curve in Fig. 15, the smallest beam diameter value presents the spot size, x-

coordinate value at the pole point indicates the working distance, and the distance range of

the curve indicates the depth of field. The working distance of 0.18 mm, depth of field of

0.16 mm, and spot size of 13 m were obtained for the sample #1 with 0.60 mm length of

100/140 GRIN fiber lens which was directly attached to SMF without fiber spacer. If the

length of GRIN fiber was reduced to 0.52 mm, the beam became less converged resulting the

(a)

(b)

three parameters increased slightly to 0.28 mm, 0.5 mm, and 22 m, respectively. When a

Fig. 16. In vivo human finger SS-OCT images taken with probe #14 (working distance, depth

0.48 mm fiber spacer was inserted between SMF and GRIN fiber and the length of GRIN

of field and spot diameter of 1.2 mm 1.1 mm, and 27 m) and # 16 (working distance, depth

fiber is reduced to 0.17 mm, significant less converged beam is observed. The working

of field and spot diameter of 0.9 mm 0.33 mm, and 16 m) acquired by our catheter-based

distance of 1.0 mm, depth of field of 0.95 mm, and spot size of 28 m were obtained for

complex SS-OCT using our 3x3 Mach-Zehnder interferometer with unbalanced differential

sample #8.

detection technique with image size of 5x2mm.

The measured beam diameters are well matched to Gaussian-fitted values in the center

Fig. 17 shows ex vivo images of coronary artery of rabbit with forward-view (a, b) probe with

(focused) regions, but have small deviations in the two edge side regions as shown in Fig. 15.

ball lens #16 and side-view (c, d) probe with GRIN lens #5 acquired by our catheter-based

The insets in Fig. 15 shown images of the cross-section beam profile at begin, center, and

complex SS-OCT using our 3x3 Mach-Zehnder interferometer with unbalanced differential

end planes, respectively, for the sample #8. The circular shapes of the profile images

detection technique with image size of 2.5x2mm by scanning the probe along the artery (a, c)

indicate high x and y symmetry of the beam profiles through all the range of depth of filed.

and scanning cross the artery (b, d). Three layers of tunica intima, tunica media, and tunica

adventitia for the coronary artery are viewed clearly as indicated by the gray, black, and

Considering chromatic aberrations, from ZEMAX simulation for the ball fiber lens in the

white arrows in all four images in Fig. 17. The fine layers of muscle and elastic fiber in the

wavelength range of 1260 – 1370 nm, the relative variations of the working distance, depth

tunica media of the coronary artery are shown obviously in the images obtained by

of field and spot size were calculated all smaller than 4.0%. For the GRIN fiber lens, the

scanning the probe cross the artery.

range of the zero-dispersion wavelengths, is 1297-1316 nm. The zero-dispersion slope, S

0

0,

equal to or smaller than 0.101 ps/nm2-km. Using the standard formula of fiber dispersion,

D()  S   

(ps/nm-km), we calculated the changes of refractive index in the

0[

4

0 / 3 ] / 4

1260 – 1370 nm wavelength range. By using these values in ZEMAX, we calculated the

index-59_1.png

index-59_2.png

Full Range Swept-Source Optical Coherence

Tomography with Ultra Small Fiber Probes for Biomedical Imaging

51

relative changes of the working distance, depth of field and spot size were all smaller than

3%. Based on our results, the desired beam profile for the application of optical biomedical

60

imaging systems can be obtained by the GRIN and ball fiber lens with or without fiber

) 50

spacers. The technique described here possesses a high degree of flexibility for designing

ultra-small optical probes with different beam shapes for the different tissue imaging.

40

eter (m

5. OCT Imaging

30

#8

iam

D

#3

Measured X

Fig. 16 (a) and (d) show SS-OCT images of human finger in vivo taken using fiber probes #

20

Measured Y

eam

14 (working distance, depth of field and spot diameter of 1.2 mm 1.1 mm, and 27 m) and #

#1

Gaussian Fitted X

B

Gaussian Fitted Y

16 (working distance, depth of field and spot diameter of 0.9 mm 0.33 mm, and 16 m)

10

acquired by our catheter-based complex SS-OCT using our 3x3 Mach-Zehnder

0.0

0.5

1.0

1.5

2.0

interferometer with unbalanced differential detection technique with image size of 5x2mm.

Distance to Lens Surface (mm)

The image depth shown in Fig. 16 (a) is slightly larger than that in Fig. 16 (b), but the image

Fig. 15. Measured and Gaussian-fitted 1/e2 intensity beam diameters along the axial

is blurrier in Fig. 16 (a) than that in Fig. 16 (b), which taken by the probe with larger depth of

distance (zero is the position of the lens surface) at x (horizontal) and y (vertical) radial

field and spot size. The image shown in Fig. 16 (b) has higher resolution than that in Fig. 16

coordination in the distance range of depth of field of the samples #1, #3 and #8, which was

(a), which can be seen clearly with finer structures in layer of epidermis (grey arrow), sweat

made from the GRIN fiber lens. Insets: the cross-section beam profiles at three planes for the

gland (white arrow), and blood vessel in subcutis layer (black arrow).

sample #8.

For each curve in Fig. 15, the smallest beam diameter value presents the spot size, x-

coordinate value at the pole point indicates the working distance, and the distance range of

the curve indicates the depth of field. The working distance of 0.18 mm, depth of field of

0.16 mm, and spot size of 13 m were obtained for the sample #1 with 0.60 mm length of

100/140 GRIN fiber lens which was directly attached to SMF without fiber spacer. If the

length of GRIN fiber was reduced to 0.52 mm, the beam became less converged resulting the

(a)

(b)

three parameters increased slightly to 0.28 mm, 0.5 mm, and 22 m, respectively. When a

Fig. 16. In vivo human finger SS-OCT images taken with probe #14 (working distance, depth

0.48 mm fiber spacer was inserted between SMF and GRIN fiber and the length of GRIN

of field and spot diameter of 1.2 mm 1.1 mm, and 27 m) and # 16 (working distance, depth

fiber is reduced to 0.17 mm, significant less converged beam is observed. The working

of field and spot diameter of 0.9 mm 0.33 mm, and 16 m) acquired by our catheter-based

distance of 1.0 mm, depth of field of 0.95 mm, and spot size of 28 m were obtained for

complex SS-OCT using our 3x3 Mach-Zehnder interferometer with unbalanced differential

sample #8.

detection technique with image size of 5x2mm.

The measured beam diameters are well matched to Gaussian-fitted values in the center

Fig. 17 shows ex vivo images of coronary artery of rabbit with forward-view (a, b) probe with

(focused) regions, but have small deviations in the two edge side regions as shown in Fig. 15.

ball lens #16 and side-view (c, d) probe with GRIN lens #5 acquired by our catheter-based

The insets in Fig. 15 shown images of the cross-section beam profile at begin, center, and

complex SS-OCT using our 3x3 Mach-Zehnder interferometer with unbalanced differential

end planes, respectively, for the sample #8. The circular shapes of the profile images

detection technique with image size of 2.5x2mm by scanning the probe along the artery (a, c)

indicate high x and y symmetry of the beam profiles through all the range of depth of filed.

and scanning cross the artery (b, d). Three layers of tunica intima, tunica media, and tunica

adventitia for the coronary artery are viewed clearly as indicated by the gray, black, and

Considering chromatic aberrations, from ZEMAX simulation for the ball fiber lens in the

white arrows in all four images in Fig. 17. The fine layers of muscle and elastic fiber in the

wavelength range of 1260 – 1370 nm, the relative variations of the working distance, depth

tunica media of the coronary artery are shown obviously in the images obtained by

of field and spot size were calculated all smaller than 4.0%. For the GRIN fiber lens, the

scanning the probe cross the artery.

range of the zero-dispersion wavelengths, is 1297-1316 nm. The zero-dispersion slope, S

0

0,

equal to or smaller than 0.101 ps/nm2-km. Using the standard formula of fiber dispersion,

D()  S   

(ps/nm-km), we calculated the changes of refractive index in the

0[

4

0 / 3 ] / 4

1260 – 1370 nm wavelength range. By using these values in ZEMAX, we calculated the

index-60_1.png

index-60_2.png

index-60_3.png

index-60_4.png

index-60_5.png

index-60_6.png

index-60_7.png

index-60_8.png

index-60_9.png

index-60_10.png

52

Biomedical Imaging

(b)

(c)

(d)

6. Acknowledgments

Authors gratefully thank Dan P. Popescu and Michael G. Sowa from Institute for

Biodiagnostics of National Research Council of Canada, and Tim Cheung from Heart

Institute of University of Ottawa for their providing of the animal tissues.

(a)

(b)

(c)

(d)

Fig. 17. Ex vivo images of coronary artery of rabbit scanning along the artery (a, c) and

7. Conclusion

scanning cross the artery (b, d) with forward-view probe (a, b) with ball lens #16 and side-

view (c, d) probe with GRIN lens #5 acquired by our catheter-based complex SS-OCT using

We have demonstrated a catheter-based full range swept-source optical coherence

our 3x3 Mach-Zehnder interferometer with unbalanced differential detection technique.

tomography system. A 3x3 quadrature Mach-Zehnder interferometer with a new

unbalanced differential detection method for SS-OCT have been firstly presented. Using this

Fig. 18 shows ex vivo images of heart atrium (a, b), artery (c) of mice, and heart atrium (d, e, f)

setup, a complex conjugate artifact suppression of 27 dB has been achieved. A 90o phase

of pig with forward-view probe with ball lens #16 acquired by our catheter-based complex

shift between the two interferometric outputs was obtained thereby eliminating the need for

SS-OCT using our 3x3 Mach-Zehnder interferometer with unbalanced differential detection

further trigonometric calculations. Also, our setup resulted in a 4 dB increase in the signal-

technique with image size of 2x2mm. The shapes and structures of these internal organs of

to-noise ratio compared to a similar interferometer with the commonly used balanced

small and large animals are viewed clearly from these OCT images.

detection technique. We have then demonstrated a high-performance wavelength-swept

laser that uses a high-efficiency semiconductor optical amplifier, a high-speed polygon-

based narrowband scanning filter, and a Fourier domain mode lock technique. This laser

produced 71 mW average output power with an instantaneous linewidth of 0.09 nm, and it

could be tuned over a wavelength range of 113 nm at a repetition rate of 65.95 kHz. We also

constructed an OCT system that uses our laser source where we have shown that its

parameters are optimized for this application. We presented next the design, construction

and beam profile characterization of different variations of graded-index and ball fiber

lenses, which were recently proposed for ultra-small probes for OCT imaging. Those fiber

(a)

(b)

(c)

lens modules were made of single mode fibers and GRIN and ball fiber lenses with/without

fiber spacers between them. We used fusion-splicing in between the fibers, lenses and

spacers to ensure high quality light transmission. We found that beam-distance profiles (i.e.

0.4 - 1.2 mm of focus distance, 0.8 – 1.5 mm of depth of field, and 26 – 35 m of spot size) can

be obtained by precisely adjusting the lengths of the fiber spacer and the GRIN fiber lens or

diameter of the ball lens for the different tissue imaging in human beings and animals. We

obtained very high quality focused Gaussian beam profiles with high x and y symmetry

using the conventional multi-mode GRIN fibers and home-made fiber ball lenses. Their high

quality beam and ultra-small size make such fiber lens based probes very valuable for

(d)

(e)

(f)

biomedical optical imaging systems. The in vivo and ex vivo OCT images acquired by our

Fig. 18. Ex vivo images of heart atrium (a, b), artery (c) of mice, and heart atrium (d, e, f) of

catheter-based full range SS-OCT system indicate that this system is capable of imaging the

pig acquired by our catheter-based complex SS-OCT using our 3x3 Mach-Zehnder

biomedical tissues or inside organs for human, small/big animals and it is most suitable for

interferometer with unbalanced differential detection technique.

applications of diagnosis and guided surgery.

The in vivo and ex vivo OCT images shown in Fig. 16, 17, and 18 indicate that our catheter-

based complex SS-OCT system is capable of imaging the biomedical tissues or the inside

8. References

organs for human, small and big animals and it is most suitable for applications of diagnosis

Bilenca, A. et al. (2006). Numerical study of wavelength-swept semiconductor ring lasers:

and guided surgery.

the role of refractive-index nonlinearities in semiconductor optical amplifier and

implications for biomedical imaging application, Opt. Lett. 31, 760-762, ISSN: 0146-

9592.

Chinn, S.R.; Swanson, E. A. & Fujimoto, J. E. (1997). Optical coherence tomography using a

frequency-tunable optical source, Opt. Lett. , 22, 340-342, ISSN: 0146-9592.

Full Range Swept-Source Optical Coherence

Tomography with Ultra Small Fiber Probes for Biomedical Imaging

53

(b)

(c)

(d)

6. Acknowledgments

Authors gratefully thank Dan P. Popescu and Michael G. Sowa from Institute for

Biodiagnostics of National Research Council of Canada, and Tim Cheung from Heart

Institute of University of Ottawa for their providing of the animal tissues.

(a)

(b)

(c)

(d)

Fig. 17. Ex vivo images of coronary artery of rabbit scanning along the artery (a, c) and

7. Conclusion

scanning cross the artery (b, d) with forward-view probe (a, b) with ball lens #16 and side-

view (c, d) probe with GRIN lens #5 acquired by our catheter-based complex SS-OCT using

We have demonstrated a catheter-based full range swept-source optical coherence

our 3x3 Mach-Zehnder interferometer with unbalanced differential detection technique.

tomography system. A 3x3 quadrature Mach-Zehnder interferometer with a new

unbalanced differential detection method for SS-OCT have been firstly presented. Using this

Fig. 18 shows ex vivo images of heart atrium (a, b), artery (c) of mice, and heart atrium (d, e, f)

setup, a complex conjugate artifact suppression of 27 dB has been achieved. A 90o phase

of pig with forward-view probe with ball lens #16 acquired by our catheter-based complex

shift between the two interferometric outputs was obtained thereby eliminating the need for

SS-OCT using our 3x3 Mach-Zehnder interferometer with unbalanced differential detection

further trigonometric calculations. Also, our setup resulted in a 4 dB increase in the signal-

technique with image size of 2x2mm. The shapes and structures of these internal organs of

to-noise ratio compared to a similar interferometer with the commonly used balanced

small and large animals are viewed clearly from these OCT images.

detection technique. We have then demonstrated a high-performance wavelength-swept

laser that uses a high-efficiency semiconductor optical amplifier, a high-speed polygon-

based narrowband scanning filter, and a Fourier domain mode lock technique. This laser

produced 71 mW average output power with an instantaneous linewidth of 0.09 nm, and it

could be tuned over a wavelength range of 113 nm at a repetition rate of 65.95 kHz. We also

constructed an OCT system that uses our laser source where we have shown that its

parameters are optimized for this application. We presented next the design, construction

and beam profile characterization of different variations of graded-index and ball fiber

lenses, which were recently proposed for ultra-small probes for OCT imaging. Those fiber

(a)

(b)

(c)

lens modules were made of single mode fibers and GRIN and ball fiber lenses with/without

fiber spacers between them. We used fusion-splicing in between the fibers, lenses and

spacers to ensure high quality light transmission. We found that beam-distance profiles (i.e.

0.4 - 1.2 mm of focus distance, 0.8 – 1.5 mm of depth of field, and 26 – 35 m of spot size) can

be obtained by precisely adjusting the lengths of the fiber spacer and the GRIN fiber lens or

diameter of the ball lens for the different tissue imaging in human beings and animals. We

obtained very high quality focused Gaussian beam profiles with high x and y symmetry

using the conventional multi-mode GRIN fibers and home-made fiber ball lenses. Their high

quality beam and ultra-small size make such fiber lens based probes very valuable for

(d)

(e)

(f)

biomedical optical imaging systems. The in vivo and ex vivo OCT images acquired by our

Fig. 18. Ex vivo images of heart atrium (a, b), artery (c) of mice, and heart atrium (d, e, f) of

catheter-based full range SS-OCT system indicate that this system is capable of imaging the

pig acquired by our catheter-based complex SS-OCT using our 3x3 Mach-Zehnder

biomedical tissues or inside organs for human, small/big animals and it is most suitable for

interferometer with unbalanced differential detection technique.

applications of diagnosis and guided surgery.

The in vivo and ex vivo OCT images shown in Fig. 16, 17, and 18 indicate that our catheter-

based complex SS-OCT system is capable of imaging the biomedical tissues or the inside

8. References

organs for human, small and big animals and it is most suitable for applica